Handheld XRF: How it works

When your business considers an XRF purchase, there are many details to consider and many questions you may have. What is XRF? What does XRF do? What elements does it analyze? Is XRF accurate? Is it fast? If you and your coworkers are asking yourselves these questions, you will find some helpful answers below.

How Handheld XRF Works: A Step-by-Step Guide

XRF是X射线荧光的首字母缩写词,该过程使电子从其原子轨道位置移动,从而释放出特定元素的特征的能量爆发。然后,检测器在当时的XRF仪器中注册了该能量的释放,进而将能量按元素分类。这是该过程的详细分解:

  • 具有足够能量的X射线光束可以影响样品中原子内壳中的电子,由手持分析仪内部的X射线管产生。然后从手持XRF分析仪的前端发射X射线梁。
  • The x-ray beam then interacts with the atoms in the sample by displacing electrons from the inner orbital shells of the atom. This displacement occurs as a result of the difference in energy between the primary x-ray beam emitted from the analyzer and the binding energy that holds electrons in their proper orbits; the displacement happens when the x-ray beam energy is higher than the binding energy of the electrons with which it interacts. Electrons are fixed at specific energies in their positions in an atom, and this determines their orbits. Additionally, the spacing between the orbital shells of an atom is unique to the atoms of each element, so an atom of potassium (K) has different spacing between its electron shells than an atom of gold (Au), or silver (Ag), etc.
  • When electrons are knocked out of their orbit, they leave behind vacancies, making the atom unstable. The atom must immediately correct the instability by filling the vacancies that the displaced electrons left behind. Those vacancies can be filled from higher orbits that move down to a lower orbit where a vacancy exits. For example, if an electron is displaced from the innermost shell of the atom (the one closest to the nucleus), an electron from the next shell up can move down to fill the vacancy. This is fluorescence.
  • Electrons have higher binding energies the further they are from the nucleus of the atom. Therefore, an electron loses some energy when it drops from a higher electron shell to an electron shell closer to the nucleus. The amount of energy lost is equivalent to the difference in energy between the two electron shells, which is determined by the distance between them. The distance between the two orbital shells is unique to each element, as mentioned above.
  • The energy lost can be used to identify the element from which it emanates, because the amount of energy lost in the fluorescence process is unique to each element. The individual fluorescent energies detected are specific to the elements that are present in the sample. In order to determine the quantity of each element present, the proportion in which the individual energies appear can be calculated by the instrument or by other software.

The entire fluorescence process occurs in small factions of a second. A measurement using this process and a modern handheld XRF gun can be made in a matter of seconds. The actual time required for a measurement will depend on the nature of the sample and the levels of interest. High percentage levels will take a few seconds while part-per-million levels will take a few minutes.